Precise particle size analysis for dry products in process environments and automated laboratories from 0.25 µm to 3,500 µm
MYTOS integrates the established core technologies of HELOS laser diffraction and RODOS dry dispersion in a single robust body: a process-proven laser diffraction sensor for dry powders ranging from 0.25 µm to 3,500 µm. In combination with continuously operating sampling system TWISTER a complete integration into the processing pipe is realised making it a true in-line solution delivering representative results. Alternatively, MYTOS may be operated on-line either with TWISTER, MIXER, SCREWSAMPLER, L-probes or existing sampling solutions. Integrated at-line operation in automated lab environments is succeeded with vibratory feeder VIBRI. For instance, typical applications comprise milling operations or spray granulation.
The process-specific laser diffraction sensors find a wide range of applications from cement to pharmaceuticals, from roughest industrial conditions to GMP environments. If required, both systems may be configured for ATEX zones.
MYTOS & Co. Family
The MYTOS family includes different solutions according to their installation environment:
Sensor | Versions | Measuring range/ Modules |
---|---|---|
in-line | MYTOS & TWISTER | 0.25 µm to 1,750 µm/ 5 |
on-line | MYTOS & TWISTER | 0.25 µm to 3,500 µm/ 6 |
on-line | MYTOS & MIXER | 0.25 µm to 1,750 µm/ 6 |
on-line | MYTOS & SCREWSAMPLER | 0.25 µm to 1,750 µm/ 5 |
on-line | MYTOS & Probes | 0.25 µm to 3,500 µm/ 6 |
at-line | MYTOS & VIBRI | 0.25 µm to 3,500 µm/ 7 |
at-line | MYTOS & VIBRI Module | 0.25 µm to 1,750 µm/ 5 |
off-line | MYTOS & VIBRI Module | 0.25 µm to 1,750 µm/ 5 |
- Efficient dry dispersion ensuring complete dispersion for fine, even cohesive powders down to the submicron range with MYTOS
- Real-time particle size analysis with laser diffraction of perfectly dispersed powders with up to 2,000 measurements per second
- Representative sampling in or out of process pipes in the diameter range from 38 to 660 mm at product temperatures of up to 150°C with TWISTER
- Comparability to the established off-line laser diffraction system HELOS | RODOS that share the same core components for a valid method transfer
- Connection to process control systems | TCP/IP, OPC, Modbus®, Profibus®, PLC signals
Are you interested in additional content? Please register an account. After confirming the registration link brochures, application notes and other documents on particle measurement will be available for download.
The laser diffraction sensors MYTOS form together with representative samplers like TWISTER, MIXER, SCREWSAMPLER and probes perfectly coordinated and robust on-line process analysis systems. The sampler is integrated in the process line, takes a sample and feeds it directly to the particle size analysis.
Configurations
The at-line versions of the robust dispersion and sensor units MYTOS combined with the dosing unit VIBRI are suitable for automated laboratory operations. The integration in the central control system and the reporting of results are adapted to the existing communication environment. The system is either mounted stationary to a wall or frame or in a mobile rack.